Tecan Cavro Syringe Pump — MCP Server
Let an AI agent initialize, aspirate and dispense microlitre volumes at set flow rates, switch valve ports and stop Tecan Cavro OEM syringe pumps through the documented Cavro Data Terminal (DT) protocol.
Try it without hardware
uvx labmcp-cavro --simulate --check
The simulator is an XLP 6000 with a 1 mL syringe and a 3-port valve. Moves take real time; add --option sim_speed=10 to speed it up.
Connect your pump
Add to your MCP client
Claude Code
claude mcp add cavro -- uvx labmcp-cavro --address /dev/ttyUSB0 --option syringe_ul=1000 --option model=xcalibur
Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)
{
"mcpServers": {
"cavro": {
"command": "uvx",
"args": ["labmcp-cavro", "--address", "/dev/ttyUSB0",
"--option", "syringe_ul=1000", "--option", "model=xcalibur"]
}
}
}
Add --read-only to allow status reads but block every move.
Tools
aspirate_ul | ⚠️ hazard | Draw volume_ul into the syringe at flow_ul_s through the current (or given) valve port, and wait until the move has finished. The pump must be initialized and have room for the volume. |
dispense_ul | ⚠️ hazard | Push volume_ul out of the syringe at flow_ul_s through the current (or given) valve port, and wait until the move has finished. The syringe must contain at least that volume. |
get_command_log | 👁 read | Return the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done. |
get_connection_info | 👁 read | Report which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first. |
get_status | 👁 read | Report whether the pump is ready or busy, any error (decoded), the plunger position as the volume in the syringe, the valve position, resolution mode and current top speed. |
initialize | ⚠️ hazard | Initialize the pump: drive the plunger to the top of the syringe (expelling its contents through the valve), set that as position 0 and home the valve. Needed after power-up, a plunger overload or terminate. Route the valve output to waste first. Takes a few seconds. |
reconnect | 🛑 safety | Close and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged). |
set_valve | ⚠️ hazard | Turn the valve to a named position (3/4-port valves) or to a numbered port (distribution valves). In bypass the plunger cannot move. Returns the new status. |
terminate | 🛑 safety | Stop any plunger move, loop or delay immediately (DT command T). A valve move in progress still completes. Re-initialize afterwards: the plunger may have lost steps. |
Safety limits
max_volume_ul | 5000 µL | Largest volume one aspirate_ul/dispense_ul may move |
max_flow_ul_s | 500 µL/s | Highest plunger flow rate an agent may use |
Override at launch: --limit max_flow_ul_s=50. Independently of the limits, the server refuses moves that would overfill or over-empty the syringe, and flows the pump cannot produce with the installed syringe (top speed 5–6000 pulses/s).
Example prompts
- "Initialize the pump, then prime it: aspirate 1 mL from the input port and dispense it to the output port, three times."
- "Aspirate 150 µL of sample slowly (20 µL/s) from port 3 and dispense 50 µL into port 5."
- "How much liquid is in the syringe right now and which port is the valve on?"
- "The pump reported a plunger overload. What does that mean and what should I check?"
- "Stop the pump now."
Notes
- Conversion. Steps = volume × steps-per-stroke ÷ syringe volume, where steps-per-stroke is 6000 (XLP/XMP) or 3000 (XCalibur) in standard mode and 8× that in fine-positioning mode (N1). The mode is read from the pump (
?28) before every move. Top speed V = flow × 6000 ÷ syringe volume: on all three models a full stroke is 6000 speed pulses in N0/N1. For the XCalibur this is derived from Tecan's published 1.2 s–20 min per stroke, not from its operating manual; please confirm on hardware.
- Busy/ready is read only from
Q, as the manual requires. Invalid commands are reported immediately; invalid operands (e.g. a move past the end of the syringe) are reported by the next Q, and the server turns both into clear errors.
- Initialization force is chosen from the syringe size (full ≥ 1 mL, half for 250/500 µL, third for 50/100 µL; manual table 3-6) unless you pick one.
terminate sends T without R: in the DT command set R resumes a terminated string. T does not stop a valve move in progress. Re-initialize after terminating. If a move or initialization is still busy well past its expected duration (1.5× + 10 s), or the link fails while waiting, the server sends T itself instead of leaving the plunger moving.
- The server does not use microstep mode (N2), the OEM protocol (checksums, sequence numbers) or CAN.
- Cavro Centris has a different resolution (181,490 increments per stroke) and volume-based firmware commands that we could not verify, so it is not supported. XE 1000 and older XL/XP 3000 pumps may work with
--option steps_per_stroke=…, but they are untested.
Hardware verification